Direct Maskess Aligner und Laserlithographiegeräte Referenznummer der Bekanntmachung: WSI/2022-E24/2
Auftragsbekanntmachung
Lieferauftrag
Abschnitt I: Öffentlicher Auftraggeber
Nationale Identifikationsnummer: DE811193231
Postanschrift:[gelöscht]
Ort: Garching
NUTS-Code: DE212 München, Kreisfreie Stadt
Postleitzahl: 85748
Land: Deutschland
Kontaktstelle(n):[gelöscht]
E-Mail: [gelöscht]
Telefon: [gelöscht]
Internet-Adresse(n):
Hauptadresse: www.wsi.tum.de
Adresse des Beschafferprofils: www.wsi.tum.de
Abschnitt II: Gegenstand
Direct Maskess Aligner und Laserlithographiegeräte
Direct Maskess Aligner and Laser Lithoghraphy tools to be installed in the laboratories of the Technical University of Munich. A translation of the entire document into German language is available on request.
The desktop and stand-alone maskless laser writers shall be used for contact-free lithography on hybrid (superconducting and semiconducting) quantum circuits and devices, nanofabrication without the need for conventional lithography masks and avoiding all physical contact with delicate quantum devices. Hereby, two instruments that will be used in tandem are required: (i) a table-top system that will be used in combination with a thermolithography system for rapid prototyping and quantum device imaging and repair and provide a resolution of ~1µm and (ii) a stand-alone system that can be used for higher resolution (<0.8µm) rapid, large scale (15cm x 15cm) patterning of large arrays of devices. Both devices shall consist of direct laser lithography units, imaging and autofocus systems that will allow for precise positioning and stitching of multiple lithographic write fields and overlays. The table-top device will be operated within an ISO-4 area of a clean room, while the stand-alone device should be incorporated within an environmental chamber with a filter system providing a stable environment for the system. The system shall be equipped with all components necessary for operation, including controllers, an acoustic/vibration isolation enclosure (if necessary), computer and software, scanning optics. Freight, warranty, installation, and training costs shall be included in the final total.
Additional specific performance metrics of the two systems can be found later in this tender and separately in the detailed technical description document (downloaded from the same URL on www.evergabe.de)
Garching, DE
Direct Maskess Aligner und Laserlithographiegeräte, die in den Laboren des Physik Departments der Technischen Universität München installiert werden sollen.
Direct Maskess Aligner and Laser Lithoghraphy tools to be installed in the laboratories of the Technical University of Munich. A translation of the entire document into German language is available on request.
1. General system requirements
1.1. Design of the system(s)
The desktop and stand-alone maskless laser writers shall be used for contact-free lithography on hybrid (superconducting and semiconducting) quantum circuits and devices, nanofabrication without the need for conventional lithography masks and avoiding all physical contact with delicate quantum devices. Hereby, two instruments that will be used in tandem are required: (i) a table-top system that will be used in combination with a thermolithography system for rapid prototyping and quantum device imaging and repair and provide a resolution of ~1µm and (ii) a stand-alone system that can be used for higher resolution (<0.8µm) rapid, large scale (15cm x 15cm) patterning of large arrays of devices. Both devices shall consist of direct laser lithography units, imaging and autofocus systems that will allow for precise positioning and stitching of multiple lithographic write fields and overlays. The table-top device will be operated within an ISO-4 area of a clean room, while the stand-alone device should be incorporated within an environmental chamber with a filter system providing a stable environment for the system. The system shall be equipped with all components necessary for operation, including controllers, an acoustic/vibration isolation enclosure (if necessary), computer and software, scanning optics. Freight, warranty, installation, and training costs shall be included in the final total.
1.2. Performance specifications of “table-top” system
• The system should be delivered with a passive (or better) vibration isolation system and be portable for positioning in different areas of the laboratory.
• The system should be delivered with a sample stage system that features linear motors (or equivalent), air bearings, linear encoder position control, and vacuum chuck for various substrate sizes.
• The system should be capable of accommodating maximum substrate sizes exceeding 5” x 5” (127mm x 127mm) , minimum substrate sizes below 8 mm x 8 mm and allow write areas exceeding 80 x 80 mm2.
• The system should be able to hold substrates with a thickness <200µm (minimum) and a maximum of (at least) 10mm.
• The system should feature an overview camera system for alignment to existing structures, an intensity measurement unit for testing and calibrating the dose of the laser beam, a real-time autofocus system with a dynamic range exceeding 150µm, system control electronics for encoders, stage and feature an optical autofocus system.
• The spatial resolution of the system should be at least 1µm for the slowest scan speed, increasing at the expense of resolution.
• The system should be delivered with all required control electronics, computer control and software allowing DXF and GDSII files to be read into the system and written.
• The system should be compatible with broadband and i-line photoresists, including medusa, SU-8. AZ-mOF, TOK IP.
• The control computer should be delivered with the system and offer CAD software for on the fly design of writing patterns.
1.3. Performance specifications of “stand alone” system
• The system should be entirely contained within an environmental chamber that provides temperature control, laminar flow of clean, filtered air (ISO 5 or better) and a stable operating environment. This environment should provide > 0.25 m / s laminar flow , adjustable up to 0.5 m/s. The temperature stability should be + / - 1C.
• The system should be delivered with a passive (or better) vibration isolation system and be portable for positioning in different areas of the laboratory.
• The system should be delivered with a sample stage system that features linear motors (or equivalent), air bearings, linear encoder position control, and vacuum chuck for various substrate sizes.
• The system should be capable of accommodating maximum substrate sizes exceeding 5” x 5” (127mm x 127mm) , minimum substrate sizes ideally below 3 mm x 3 mm and allow write areas exceeding 80 x 80 mm2.
• The system should be able to hold substrates with a thickness <200µm (minimum) and a maximum of (at least) 10mm.
• The system should feature an overview camera system for coarse alignment to existing structures, an intensity measurement unit for testing and calibrating the dose of the laser beam, a real-time autofocus system with a dynamic range exceeding 150µm, system control electronics for encoders, stage and feature an optical autofocus system.
• The system should feature an interferometer that facilitates lateral movement of the substrate with a relative positioning accuracy <20nm.
• The spatial resolution of the system should be at least 0.7µm.
• The system should be delivered with all required control electronics, computer control and software allowing DXF and GDSII files to be read into the system and written.
• The system should feature the possibility to align exposures to structure on the backside of the substrate using a camera under the substrate or similar with an alignment precision of at least 3µm.
• The system should be capable of exposing a 4” substrate in < 2 hours.
• The system should be compatible with broadband and i-line photoresists, including medusa, SU-8. AZ-mOF, TOK IP.
• The control computer should be delivered with the system and offer CAD software for on-the-fly design of writing patterns.
2. Delivery and acceptance test
2.1. Delivery date, installation & training
The system must be delivered within 6 months after receipt of the order and pass the acceptance test at the TUM.
The total price must include delivery / freight, installation costs, VAT, training costs, and warranty.
At least 2 employees of the TUM should be trained in the use of each of the systems.
2.2. Warranty
The system must include a warranty of 12 months from successful completion of the acceptance test against all faults and failures provided by the supplier or manufacturer.
2.3. Acceptance test
Following installation at the target laboratories of the TUM, both systems shall be tested by a technician from the supplier accompanied by an employee of the TUM. In the test, all listed specifications and features shall be fulfilled and demonstrated.
Abschnitt III: Rechtliche, wirtschaftliche, finanzielle und technische Angaben
Mitteilung von Unklarheiten in den Vergabeunterlagen Enthalten die Vergabeunterlagen nach Auffassung des Bewerbers Unklarheiten, so hat er unverzüglich die Vergabestelle vor Angebotsabgabe in Textform darauf hinzuweisen.
Mitteilung von Unklarheiten in den Vergabeunterlagen Enthalten die Vergabeunterlagen nach Auffassung des Bewerbers Unklarheiten, so hat er unverzüglich die Vergabestelle vor Angebotsabgabe in Textform darauf hinzuweisen.
Die Auswahlkriterien sind der Preis (20%), die Übereinstimmung mit möglichst vielen technischen Spezifikationen (75%) und die Lieferzeit (5%).
Bieter sollten das Formular "04 EU_Eigenerklärung Eigtnung VgV.docx" ausfüllen und dem Angebot beigefügt werden. Dieses Formular steht auch unter der gleichen URL wie die technische Beschreibung zum Download bereit (www.evergabe.de)
Der Lieferant muss in der Lage sein, Kundensupport und Service mit einer Reaktionszeit von weniger als 10 Arbeitstagen für das Gerät zu bieten.
Der Lieferant muss in der Lage sein, nach Ablauf der Garantie Ersatzteile zu liefern.
Weitere Kriterien sind in dem Dokument 02 "EU_Bewerbungsbedingungen_Europa.docx" aufgeführt, das du mit den Ausschreibungsunterlagen herunterladen kannst.
Das im Download enthaltene Formular "Eigenerklärung zur Eignung - VgV" muss mit dem Angebot eingereicht werden.
Der Lieferant muss in der Lage sein, Kundensupport und Service mit einer Reaktionszeit von weniger als 10 Arbeitstagen für das Gerät zu bieten.
Der Lieferant muss in der Lage sein, nach Ablauf der Garantie Ersatzteile zu liefern.
Abschnitt IV: Verfahren
Garching, LS Prof Finley (S209)
Prof. J. Finley, Dr. N. Wilson oder Dr. A. Stier werden alle Angebote öffnen und an Prof. S. Fillip weiterleiten. Die Gruppe wird dann über das passendste Angebot diskutieren und noch am selben Tag die Absicht verkünden, den Zuschlag zu erteilen.
Abschnitt VI: Weitere Angaben
Postanschrift:[gelöscht]
Ort: München
Postleitzahl: 80538
Land: Deutschland
E-Mail: [gelöscht]
Telefon: [gelöscht]
Fax: [gelöscht]
Internet-Adresse: www.regierung-oberbayern.de